Pre-analytical errors in Clinical Chemistry Laboratory of a Tertiary care Hospital
Figure 1. Frequency of occurrence of pre-analytical errors
Thepercentageoccurrenceofthepreanalyticalvariableswasalsodeterminedwhich
indicated that hemolysis was seen in 29.99% samples, clotted blood in 18.01% of
the samples. Collection of blood samples in wrong tube, wrong preservative and
obtaining wrong volume collectively accounted for faulty results in 31.02% of the
samples due to which they were rejected. Incorrect timing of collection of samples
was seen in 9% of the total samples. Mislabelling of the samples by the laboratory
personnel was seen as a cause of rejection of 7.06% of the samples. 4.92% lipemic
(milky) samples were rejected being a interfering factor in analysis (Figure 2).
Overall prevalence of pre analytical errors in our laboratory was found
to be 2.32%
By far the most prevalent pre analytical variable seen responsible for
sample rejection was found to be hemolysis seen in 29.99% of samples.
Closely followed by the addition of clotted blood as the second most
common factor responsible for sample rejection was seen in 18.01% of
the samples.